A bone curette is a hand instrument with a sharp-edged cup on the end of a shaft, used to scrape and scoop bone, granulation tissue and debris. Volkmann, Bruns, Spratt and the other named patterns mostly differ in overall length, cup shape, single or double ends and whether the shank is straight or angled. Within each pattern, the "Fig." number is the cup-size step.
Quick answer Pick the reach first (about 13–17 cm for shallow, open work; 21–28 cm for deeper cavities), then straight or angled, then the cup size. Fig. sizes are a step within each pattern and are not standardized between makers, so match against a set you already use or ask us for the cup dimension.
What is a bone curette used for?
Quick answer Scraping. The sharpened rim of the cup lifts away soft tissue, cancellous bone or debris, and the cup holds what it removes.
Two common orthopedic uses show up clearly in the literature:
- Clearing out a bone lesion. A published technique for benign cystic bone lesions tells surgeons to "use curets, rongeurs, or spoons to remove all contents from the cavity," and to confirm the whole cavity can be reached with the curet[1]. For giant cell tumor of bone, StatPearls lists curettage, curettage with bone grafting and curettage with bone cement among the treatment options[2].
- Harvesting cancellous bone graft. In a 2023 cadaveric study of anterior iliac crest harvesting, the cancellous bone was taken with "a sharp surgical curette with a cup size of 8 mm"[3]. The same study found a trephine collected more bone volume than the curette technique, so the curette is one option among several[3].
Which technique and instrument to use is a clinical decision for the surgeon. This guide only helps you identify and order the right pattern and size.
How do Volkmann, Bruns, Spratt and other patterns differ?
Quick answer By length, number of working ends, cup shape and shank angle. Compare the photo and catalog number, not just the name.
The names are traditional pattern names. The table below uses the sizes listed in our own catalog.
| Pattern | Length(s) | Ends | Options | Fig. sizes |
|---|---|---|---|---|
| Volkmann | 17 cm (6¾″), 27 cm (10¾″) | Single | Stainless or Ferrozell handle | 0000–6 |
| Volkmann double-ended | 13–20 cm (5″–8″) | Double | Style A or Style B | Sold by length and style |
| Bruns | 17 cm (6¾″), 23 cm (9″) | Single | Straight; 45° angled on the 23 cm | 0000–6 |
| Spratt | 17 cm (6¾″) | Single | One style | 0000–6 |
| Williger | 13.5 cm (5¼″), 17.5 cm (7″) | Double or single | Oval cup or round cup | 0/0–1/2 (double); 00–2 (single) |
| Cobb | 28 cm (11″) | Single | Straight or 45° angled | 000–6 |
| Simon | 24 cm (9½″) | Single | One style | 1–6 |
| Halle | 21 cm (8¼″) | Single | One style | 1–3 |



When does a double-ended curette make sense?
Quick answer When you want two cups on one instrument and fewer instruments on the tray. Single-ended curettes give you the full handle to grip.

A double-ended curette usually pairs two different cup sizes or shapes. Our double-ended options are the Williger (13.5 cm, Fig. 0/0, 0/1 and 1/2), the Volkmann double-ended (13–20 cm, Style A or B) and the Martini double-ended (14 cm or 21 cm).
What do Fig. sizes and cup sizes mean?
Quick answer "Fig." is the catalog size step for the cup within one pattern. Order by catalog number, and ask us for the cup dimension if millimetres matter.
Our Volkmann, Bruns and Spratt curettes are listed from Fig. 0000 through Fig. 6, Cobb from Fig. 000 through Fig. 6, and Simon from Fig. 1 through Fig. 6.
- Fig. sizes are not a universal standard. The same Fig. number can mean a different cup width from another maker, or even between two patterns.
- Clinical papers usually give millimetres, not Fig. numbers. The graft study above names an 8 mm cup[3]. If you need a set cup diameter, contact us with the dimension and we will confirm which catalog number is closest.
- Order by catalog number. Each size has its own MP number (for example, the Spratt Fig. 0 is MP-19247), which avoids mix-ups when reordering.
Straight, angled or reverse angled: which shank do I need?
Quick answer Straight for surfaces in line with the shaft, angled for side walls, reverse angled for undercuts you work toward yourself.
| Shank | Where it reaches | Where we offer it |
|---|---|---|
| Straight | Surfaces directly in line with the shaft | Listed as an option on Bruns, Cobb, Long Shaft and USA Pattern |
| Angled (45° or "B") | Side walls of a cavity | Bruns 23 cm, Cobb, Long Shaft, USA Pattern |
| Reverse angled ("C") | Undercuts and back walls, working toward you | Long Shaft, USA Pattern |

Deep cavities may also call for a long pattern such as the 25 cm Long Shaft curette or the 28 cm Cobb curette.
How should reusable curettes be cleaned and inspected?
Quick answer Clean first, inspect inside the cup, then sterilize, following the instrument's instructions for use and your facility's protocols.
A curette is a critical item because it enters sterile tissue. CDC guidance says objects that enter sterile tissue "must be sterile because any microbial contamination could transmit disease"[4]. The cup traps bone and tissue, so cleaning matters before any sterilization step:
- Clean first. CDC states that "thorough cleaning is required before high-level disinfection and sterilization," and advises presoaking or rinsing surgical instruments so blood does not dry on them[5].
- Inspect every instrument. CDC says all instruments "should be individually inspected and be visibly clean"[5]. On a curette, look inside the cup and along the rim.
- Follow the labeling. FDA describes reprocessing as a multistep process of cleaning, then disinfection or sterilization, following the device's labeling instructions[6].
For more detail, see our guide on caring for reusable stainless steel instruments.
Featured in this guide
Volkmann Bone Curette17 or 27 cm · Fig. 0000–6
Bruns Bone CuretteStraight or 45° · Fig. 0000–6
Spratt Bone Curette17 cm · Fig. 0000–6
Williger Bone CuretteSingle or double-ended
Bone Curette, USA PatternStraight, angled or reverse · Fig. 1–6
Halle Bone Curette21 cm · Fig. 1–3
Browse the full range in Bone Curettes, or see Bone Rongeurs for the instruments often used alongside them.
FAQ
Is a Volkmann curette the same as a Bruns or Spratt curette?
No. They are separate patterns. They do the same basic job of scraping bone with a cup, but differ in length, size range and shank options. Compare the product photos and catalog numbers rather than the name alone.
Does a higher Fig. number mean a bigger cup?
Fig. numbers are a size step within one pattern, but they are not standardized across manufacturers. If cup diameter matters, ask us for the dimension before ordering.
Should I choose single-ended or double-ended?
Double-ended curettes give two working ends on one instrument and can reduce tray count. Single-ended curettes give you the whole handle to grip.
What is a Ferrozell handle?
Our Volkmann curette is offered with either an all-stainless handle or a Ferrozell handle. Check the instructions for use for any reprocessing limits that apply to the handle material.
Do I need a straight, angled or reverse-angled curette?
A straight shank reaches surfaces directly in line with the shaft, an angled shank reaches the side walls of a cavity, and a reverse-angled shank reaches undercuts and back walls, working toward you.
Sources
- JBJS Essential Surgical Techniques – Ibe IK, Alder KD, Henderson SE, Yu KE, Lee FY. Excisional Curettage of Benign Cystic Lesions (2019). PMC6948994
- NCBI Bookshelf / StatPearls – Giant Cell Tumor (Osteoclastoma) (last updated 2024). NBK559229
- Chinese Journal of Traumatology – Cakar A, et al. Comparison of curettage vs. trephination technique for harvesting anterior iliac crest bone graft: A cadaveric study (2023). PMC11973664
- Centers for Disease Control and Prevention – A Rational Approach to Disinfection and Sterilization, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008). cdc.gov
- Centers for Disease Control and Prevention – Cleaning, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008). cdc.gov
- U.S. Food and Drug Administration – Reprocessing of Reusable Medical Devices (2023). fda.gov